54 Model IFT9701 Transmitter Instruction Manual
Configuration with ProLink II Software continued
Use a mass flow unit during the calibration procedure. If the application requires volume flow
measurement, choose a mass flow unit for the flow calibration, then choose a volume flow unit for the
application.
To calculate the flow calibration factor:
1. Configure the transmitter to use a mass flow measurement unit (see Section 7.2.2).
2. In the Flow panel (depicted in Figure 7-2), enter the flowmeter’s 8-digit flow calibration factor
into the Flow Cal text box. The flow calibration factor can be found on the flowmeter’s serial
number tag, or on the certificate that is shipped with the sensor.
3. Zero the flowmeter (see Section 7.3.1).
4. With flow through the meter completely stopped, from the ProLink menu, click on Totalizer
Control. The Totalizer Control window is displayed, as depicted in Figure 7-6.
Figure 7-6 Totalizer Control window
5. Click the Reset button to reset the internal totalizer to zero.