MODEL 5081-C SECTION 2.0
INSTALLATION
TABLE 2-1. Model 5081-C Sensor Selection
2.3.4 Contacting Loops.
The Model 5081-C conductivity transmitter is designed to make accurate measurements while in contact with the process
stream. Measurements can also be tailored to high temperature and/or high pressure streams.
2.3.5 Sensor Selection.
All Rosemount Analytical contacting conductivity sensors with PT100 RTd or PT1000 RTd are compatible with the Model
5081-C transmitter. Please refer to Figures 3-5 thru 3-9 for appropriate sensor to transmitter wiring. The sensor cable
should be routed through the left inlet closest to the connector.
Choose a contacting conductivity sensor that is appropriate for your process conditions and range of conductivity meas-
urement.
Cell Constant Range
0.01 0 to 200 µS/cm
0.1 0 to 2,000 µS/cm
1.0 0 to 20,000 µS/cm