INTRODUCTION
1.3.3 O2
Paramagnetic: The determination of oxygen is based on the measurement of the magnetic
susceptibility of the sample gas. Oxygen is strongly paramagnetic, while other common gases
are not. The detector used is compact, has fast response and a wide dynamic range. The long
life cell is corrosion resistant, heated and may be easily cleaned. It has rugged self-tensioning
suspension and is of welded Non-Glued construction.
Rosemount Analytical µCEM Continuous Analyzer Transmitter 1–6