
Installation Manual 15
I/O Wiring – Model 2400S Analog Transmitters
Power Supply I/O Wiring – DP and DNI/O Wiring – ANBefore You Begin
Figure 3-5 Required external voltage versus mA load resistance
Figure 3-6 HART multidrop wiring – Internal or external power
0
4
8
12
16
20
24
28
30
0 200 400 600 800 1000 1200
If communicating with HART, a minimum of 250 Ohms is required.
Total mA load resistance (Ohms)
Minimum required external voltage (Volts)
OPERATING REGION
HART-compatible
host or controller
HART-compatible
transmitter
SMART FAMILY
™
transmitter
Note: For optimum HART communication, make sure the output loop
is single-point-grounded to an instrument-grade ground.
24 VDC loop power
supply required for
HART 4–20 mA
passive transmitters
600 Ohm maximum resistance
250 Ohm minimum resistance
Model 2400S
External power
Model 2400S
Internal power
Komentarze do niniejszej Instrukcji