Emerson IFT9701 Instrukcja Obsługi

Przeglądaj online lub pobierz Instrukcja Obsługi dla Pomiary, testowanie i kontrola Emerson IFT9701. Emerson IFT9701 Instruction manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Model IFT9701 Transmitter

Micro MotionTMMicro Motion®Model IFT9701 Transmitter with Optional DisplayInstruction ManualInstruction ManualP/N 3100572, Rev. EMarch 2004

Strona 2

2 Model IFT9701 Transmitter Instruction ManualBefore You Begin continued1.3 European installationsThis Micro Motion product complies with all applicab

Strona 3

Model IFT9701 Transmitter Instruction Manual 3Before You Begin continuedGetting Started Power Supply and Output WiringMounting the Remote TransmitterB

Strona 4

4 Model IFT9701 Transmitter Instruction Manual

Strona 5 - Contents

Model IFT9701 Transmitter Instruction Manual 5Getting Started Power Supply and Output WiringMounting the Remote TransmitterBefore You BeginChapter 2Ge

Strona 6

6 Model IFT9701 Transmitter Instruction ManualGetting Started continuedFigure 2-1 Location of approvals tag2.1.2 Orientation and mountingOrient the tr

Strona 7

Model IFT9701 Transmitter Instruction Manual 7Getting Started continuedGetting Started Power Supply and Output WiringMounting the Remote TransmitterBe

Strona 8

8 Model IFT9701 Transmitter Instruction ManualGetting Started continuedFigure 2-2 Compartment tags and lockout device2.3 Jumper settingsUnless otherwi

Strona 9 - Before You Begin

Model IFT9701 Transmitter Instruction Manual 9Getting Started continuedGetting Started Power Supply and Output WiringMounting the Remote TransmitterBe

Strona 10 - Before You Begin continued

10 Model IFT9701 Transmitter Instruction ManualGetting Started continued2.3.3 Re-installing the circuit board compartment coverTo re-install the cover

Strona 11

Getting Started Power Supply and Output WiringMounting the Remote TransmitterBefore You BeginModel IFT9701 Transmitter Instruction Manual 11Chapter 3R

Strona 13 - Getting Started

12 Model IFT9701 Transmitter Instruction ManualRemotely Mounting the Transmitter continuedFigure 3-1 Installation dimensions for remote mountingDimens

Strona 14 - Getting Started continued

Model IFT9701 Transmitter Instruction Manual 13Remotely Mounting the Transmitter continuedGetting Started Power Supply and Output WiringMounting the R

Strona 15

14 Model IFT9701 Transmitter Instruction ManualRemotely Mounting the Transmitter continued3.4 Connecting the transmitter to the sensorTo connect the t

Strona 16

Model IFT9701 Transmitter Instruction Manual 15Remotely Mounting the Transmitter continuedGetting Started Power Supply and Output WiringMounting the R

Strona 17

16 Model IFT9701 Transmitter Instruction ManualRemotely Mounting the Transmitter continued3.4.2 Guidelines for conduitIf sealed metallic conduit is in

Strona 18

Model IFT9701 Transmitter Instruction Manual 17Remotely Mounting the Transmitter continuedGetting Started Power Supply and Output WiringMounting the R

Strona 19 - Chapter 3

18 Model IFT9701 Transmitter Instruction ManualRemotely Mounting the Transmitter continuedFigure 3-8 Sensor wiring compartment and sensor wiring termi

Strona 20

Model IFT9701 Transmitter Instruction Manual 19Getting Started Power Supply and Output WiringMounting the Remote TransmitterBefore You BeginChapter 4P

Strona 21 - User-supplied

20 Model IFT9701 Transmitter Instruction ManualPower Supply and Output Wiring continued4.3 Connect power supply wiring• Figure 4-1 shows the power sup

Strona 22

Model IFT9701 Transmitter Instruction Manual 21Power Supply and Output Wiring continuedGetting Started Power Supply and Output WiringMounting the Remo

Strona 23 - Figure 3-5 Shielded cable

Micro Motion®Model IFT9701 Transmitter with Optional DisplayInstruction ManualFor online technical support, refer to the EXPERT2™ tool at www.expert2.

Strona 24

22 Model IFT9701 Transmitter Instruction ManualPower Supply and Output Wiring continued4.4.1 Milliamp outputThe 4–20 mA output is an active output tha

Strona 25

Model IFT9701 Transmitter Instruction Manual 23Power Supply and Output Wiring continuedGetting Started Power Supply and Output WiringMounting the Remo

Strona 26 - Sensor wiring

24 Model IFT9701 Transmitter Instruction ManualPower Supply and Output Wiring continued4.4.3 Communication tools connected to milliamp outputProLink I

Strona 27 - Chapter 4

Model IFT9701 Transmitter Instruction Manual 25Power Supply and Output Wiring continuedGetting Started Power Supply and Output WiringMounting the Remo

Strona 28 - 85-250 VAC 20-30 VDC

26 Model IFT9701 Transmitter Instruction ManualPower Supply and Output Wiring continuedFigure 4-6 Wiring to pulse counter with internal pull-up resist

Strona 29 - Wire size Wire length

Model IFT9701 Transmitter Instruction Manual 27Power Supply and Output Wiring continuedGetting Started Power Supply and Output WiringMounting the Remo

Strona 30

28 Model IFT9701 Transmitter Instruction Manual

Strona 31

Model IFT9701 Transmitter Instruction Manual 29Configuration with Communicator TroubleshootingConfiguration with ProLink IIFlowmeter StartupChapter 5F

Strona 32

30 Model IFT9701 Transmitter Instruction ManualFlowmeter Startup continued5.4 InitializationAfter wiring has been connected, power can be supplied to

Strona 33 - Acceptable

Model IFT9701 Transmitter Instruction Manual 31Flowmeter Startup continuedConfiguration with Communicator TroubleshootingConfiguration with ProLink II

Strona 35

32 Model IFT9701 Transmitter Instruction ManualFlowmeter Startup continued3. Fill the flowmeter completely with the process fluid under normal process

Strona 36

Model IFT9701 Transmitter Instruction Manual 33Flowmeter Startup continuedConfiguration with Communicator TroubleshootingConfiguration with ProLink II

Strona 37 - Flowmeter Startup

34 Model IFT9701 Transmitter Instruction Manual

Strona 38 - Flowmeter Startup continued

Model IFT9701 Transmitter Instruction Manual 35Configuration with Communicator TroubleshootingConfiguration with ProLink IIFlowmeter StartupChapter 6C

Strona 39

36 Model IFT9701 Transmitter Instruction ManualConfiguration with a HART Communicator continued6.2 Configuration parametersUse the basic setup menu to

Strona 40

Model IFT9701 Transmitter Instruction Manual 37Configuration with a HART Communicator continuedConfiguration with Communicator TroubleshootingConfigur

Strona 41 - 5.7 Process measurement

38 Model IFT9701 Transmitter Instruction ManualConfiguration with a HART Communicator continued7. At the detailed setup menu, choose 2 (configure fiel

Strona 42

Model IFT9701 Transmitter Instruction Manual 39Configuration with a HART Communicator continuedConfiguration with Communicator TroubleshootingConfigur

Strona 43 - Chapter 6

40 Model IFT9701 Transmitter Instruction ManualConfiguration with a HART Communicator continued6.2.6 Range values for milliamp outputRange values are

Strona 44

Model IFT9701 Transmitter Instruction Manual 41Configuration with a HART Communicator continuedConfiguration with Communicator TroubleshootingConfigur

Strona 45

Model IFT9701 Transmitter Instruction Manual iContentsChapter 1 Before You Begin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 46

42 Model IFT9701 Transmitter Instruction ManualConfiguration with a HART Communicator continuedd. When the display reads “WARN Loop should be removed

Strona 47

Model IFT9701 Transmitter Instruction Manual 43Configuration with a HART Communicator continuedConfiguration with Communicator TroubleshootingConfigur

Strona 48

44 Model IFT9701 Transmitter Instruction ManualConfiguration with a HART Communicator continued6. Multiply the meter factor from Step 5 by the first f

Strona 49

Configuration with Communicator TroubleshootingConfiguration with ProLink IIFlowmeter StartupModel IFT9701 Transmitter Instruction Manual 45Chapter 7C

Strona 50

46 Model IFT9701 Transmitter Instruction ManualConfiguration with ProLink II Software continued7.2 Configuration parametersUse the Configuration windo

Strona 51

Model IFT9701 Transmitter Instruction Manual 47Configuration with ProLink II Software continuedConfiguration with Communicator TroubleshootingConfigur

Strona 52

48 Model IFT9701 Transmitter Instruction ManualConfiguration with ProLink II Software continued4. In some sensor installations, noise caused by mechan

Strona 53 - Chapter 7

Model IFT9701 Transmitter Instruction Manual 49Configuration with ProLink II Software continuedConfiguration with Communicator TroubleshootingConfigur

Strona 54

50 Model IFT9701 Transmitter Instruction ManualConfiguration with ProLink II Software continued7.2.3 Output configurationBasic configuration of flowme

Strona 55

Model IFT9701 Transmitter Instruction Manual 51Configuration with ProLink II Software continuedConfiguration with Communicator TroubleshootingConfigur

Strona 56

ii Model IFT9701 Transmitter Instruction ManualContents continuedChapter 5 Flowmeter Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 57

52 Model IFT9701 Transmitter Instruction ManualConfiguration with ProLink II Software continuedTo perform the auto zero procedure:1. Prepare the flowm

Strona 58

Model IFT9701 Transmitter Instruction Manual 53Configuration with ProLink II Software continuedConfiguration with Communicator TroubleshootingConfigur

Strona 59

54 Model IFT9701 Transmitter Instruction ManualConfiguration with ProLink II Software continuedUse a mass flow unit during the calibration procedure.

Strona 60

Model IFT9701 Transmitter Instruction Manual 55Configuration with ProLink II Software continuedConfiguration with Communicator TroubleshootingConfigur

Strona 61

56 Model IFT9701 Transmitter Instruction ManualConfiguration with ProLink II Software continuedFigure 7-7 Configuration window – Density panel

Strona 62

Model IFT9701 Transmitter Instruction Manual 57Configuration with Communicator TroubleshootingConfiguration with ProLink IIFlowmeter StartupChapter 8T

Strona 63

58 Model IFT9701 Transmitter Instruction ManualTroubleshooting continuedFigure 8-1 Transmitter terminals and diagnostic LED8.3.2 Optional LCDExcept fo

Strona 64

Model IFT9701 Transmitter Instruction Manual 59Troubleshooting continuedConfiguration with Communicator TroubleshootingConfiguration with ProLink IIFl

Strona 65 - Troubleshooting

60 Model IFT9701 Transmitter Instruction ManualTroubleshooting continuedFigure 8-2 Power supply connections7. Make sure the power cable is not install

Strona 66 - Troubleshooting continued

Model IFT9701 Transmitter Instruction Manual 61Troubleshooting continuedConfiguration with Communicator TroubleshootingConfiguration with ProLink IIFl

Strona 67 - Optional LCD reads: Condition

Model IFT9701 Transmitter Instruction Manual iiiContents continued8.7 Slug flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 68

62 Model IFT9701 Transmitter Instruction ManualTroubleshooting continuedIf the transmitter indicates an over range or sensor failure condition, follow

Strona 69 - Circuit Wire colors

Model IFT9701 Transmitter Instruction Manual 63Troubleshooting continuedConfiguration with Communicator TroubleshootingConfiguration with ProLink IIFl

Strona 70 - Digital multimeter

64 Model IFT9701 Transmitter Instruction Manual

Strona 71

Model IFT9701 Transmitter Instruction Manual 65HART Menu Trees Return PolicyOptional DisplaySpecificationsAppendix AIFT9701 SpecificationsA.1 Performa

Strona 72

66 Model IFT9701 Transmitter Instruction ManualIFT9701 Specifications continuedA.2.2 Local display (optional)5-digit, alphanumeric, liquid crystal dis

Strona 73 - IFT9701 Specifications

Model IFT9701 Transmitter Instruction Manual 67IFT9701 Specifications continuedHART Menu Trees Return PolicyOptional DisplaySpecifications20 to 30 VDC

Strona 74

68 Model IFT9701 Transmitter Instruction ManualIFT9701 Specifications continuedA.3.3 Humidity limits5 to 95% non-condensingA.3.4 Vibration limitsMeets

Strona 75

Model IFT9701 Transmitter Instruction Manual 69IFT9701 Specifications continuedHART Menu Trees Return PolicyOptional DisplaySpecificationsA.5 Environm

Strona 76

70 Model IFT9701 Transmitter Instruction Manual

Strona 77

Model IFT9701 Transmitter Instruction Manual 71HART Menu Trees Return PolicyOptional DisplaySpecificationsAppendix BHART Communicator Menu Trees 1 Per

Strona 78

iv Model IFT9701 Transmitter Instruction Manual

Strona 79 - HART Communicator Menu Trees

72 Model IFT9701 Transmitter Instruction ManualHART Communicator Menu Trees continued1 Flow calibration factor2 Density calibration factor1 Frequency

Strona 80 - *See previous

HART Menu Trees Return PolicyOptional DisplaySpecificationsModel IFT9701 Transmitter Instruction Manual 73Appendix CInstalling the Optional DisplayTo

Strona 81 - Appendix C

74 Model IFT9701 Transmitter Instruction ManualInstalling the Optional Display continued8. Put the retaining clip, washer, and screw in place on the b

Strona 82

Model IFT9701 Transmitter Instruction Manual 75HART Menu Trees Return PolicyOptional DisplaySpecificationsAppendix DReturn PolicyD.1 General guideline

Strona 83 - Return Policy

76 Model IFT9701 Transmitter Instruction Manual

Strona 84

Model IFT9701 Transmitter Instruction Manual 77 IndexIndexAAbout this manual 1Accessibility, reliability and safety 5–7Agency approvalsapproved are

Strona 85 - Index

78 Model IFT9701 Transmitter Instruction ManualIndex continuedDefinitions 2Densitycalibration factorHART Communicator 44ProLink II software 55Diagnost

Strona 86 - Index continued

Model IFT9701 Transmitter Instruction Manual 79Index continued IndexOOrientation 6Output wiring 21–23HART Communicator 24HART network 23milliamp ou

Strona 87

80 Model IFT9701 Transmitter Instruction ManualIndex continuedUULapproved areas 69hazardous area installation 5Used equipment returns 75VVibration lim

Strona 89

Model IFT9701 Transmitter Instruction Manual 1Getting Started Power Supply and Output WiringMounting the Remote TransmitterBefore You BeginChapter 1Be

Strona 90 - *3100572*

Micro MotionTMMicro Motion Inc. USAWorldwide Headquarters7070 Winchester CircleBoulder, Colorado 80301T (303) 530-8400(800) 522-6277F (303) 530-8459Mi

Komentarze do niniejszej Instrukcji

Brak uwag