E-6 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
SET METER* FOR 300
VAC. PLACE PROBES ON
TERMINAL BLOCK J5.
METER INDICATES
CURRENT LINE SETTING.
YES
YES
NO
NO
NO
YES
DISCONNECT POWER TO
IFT. CHECK FUSES ON
POWER SUPPLY BOARD.
FUSES BLOWN IN IFT.
SET METER FOR 50 VDC.
CHECK VOLTAGE TEST
POINTS ON
MICROPROCESSOR BOARD.
METER INDICATES
CORRECT VOLTAGES FOR
TEST POINTS.
REPLACE
MICROPROCESSOR
BOARD.
EXAMINE MAIN LINE
CURRENT.
CHECK THAT LINE
VOLTAGE JUMPERS ARE
SET CORRECTLY.
REPLACE POWER
SUPPLY BOARD.
REPLACE BLOWN FUSES.
SYMPTOM - MICROPROCESSOR BOARD LED IS OFF
*SIMPSON MODEL 260 OR EQUIVALENT MULTIMETER.
34990013
Figure E-3. IFT Troubleshooting Flowchart, #1