Emerson Micro Motion F-Series Sensor Instrukcja Instalacji Strona 10

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 20
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 9
Feedthrough and extender componentsFigure 2-2:
A
B
C
D
E
F
G
H
A. Transmitter or core processor
B. Extender
C. Plastic plug
D. Plastic cap
E. O-ring
F. Clamping ring
G. Clamping screw
H. Feedthrough
2. Loosen the clamping screw and remove the clamping ring. Leave the O-ring in place
on the feedthrough.
3. Remove and recycle the plastic plug from the extender.
4. Fit the extender onto the feedthrough by carefully aligning the notches on the
bottom of the extender with the notches on the feedthrough.
5. Close the clamping ring and tighten the clamping screw to 13–18 in-lbs.
(1,5–2 N-m).