Model 499A CL-01
Free Chlorine Sensor
Instruction Sheet
PN 51A-499ACL01/rev.K
September 2010
SPECIFICATIONS - SENSOR
Pressure: 0 to 65 psig (0 to 549 kPa abs)
Temperature: 32 to 122°F (0 to 50°C)
Process Connection: 1 inch MNPT
Wetted Parts: Noryl
®
1
, Viton
®
2
, platinum, polyethersulfone,
polyester, wood, silicone
Type PN Wetted Materials Process Connection Max.Temperature Max. Pressure
1-1/2 in. tee 23567-00 CPVC, Buna N 1-1/2 in. socket
122°F (50°C)
65 psig (549 kPa abs)
915240-03 3/4 in. NFPT
2 in. tee 915240-04 PVC, Buna N 1 in. NFPT 120
°
F (49
°
C) 60 psig (515 kPa abs)
915240-05 1-1/2 in. NFPT
Low Flow* 24091-00 Polycarbonate/ Compression fitting for
158°F (70°C)
90 psig (722 kPa abs)
polyester, 316SS, 1/4 in. OD tubing
silicone
24091-01 (with Polycarbonate/ Compression fitting for
158°F (70°C)
90 psig (722 kPa abs)
bubble sweep- polyester, 316SS, 1/4 in. OD tubing
ing nozzle) silicone
1
Noryl is a registered trademark of General Electric.
2
Viton is a registered trademark of DuPont Performance Elastomers
SPECIFICATIONS - FLOW CELLS
For additional information, please visit our website
at www.emersonprocess.com/raihome/liquid/.
CAUTION
Do not exceed pressure and temperature specifications.
Pressure: 65 psig max (549 kPa abs max)
Temperature: 32 to 122°F (0 to 50°C)
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible
with process com position and operating conditions.
Application compat ibility is entirely the responsibility of
the user.
*
Pressure and temperature specifications for the low flow cell exceed the temperature and pressure specifications for the sensor.
Przeglądanie stron 0
1 2 3 4 5 6 7 8

Podsumowanie treści

Strona 1 - Free Chlorine Sensor

Model 499A CL-01Free Chlorine SensorInstruction SheetPN 51A-499ACL01/rev.KSeptember 2010SPECIFICATIONS - SENSORPressure: 0 to 65 psig (0 to 549 kPa ab

Strona 2 - INSTALLATION

2MODEL 499A CL-01 INSTALLATIONFIGURE 4. Low Flow Cell (PN 24091-00 or 24091-01)INSTALLATION Install the sensor in a flowing sample. Keep the sample fl

Strona 3 - MODEL 499A CL-01 WIRING

MODEL 499A CL-01 WIRINGWIRING3FIGURE 5. Wiring Sensor with Standard Cable toModel 1056 and Model 56 AnalyzerFIGURE 6. Wiring Sensor with Optimum EMI

Strona 4

FIGURE 9. Wiring Sensor with Standard Cable toModel 5081-A Transmitter.FIGURE 10. Wiring Sensor with Optimum EMI/RFI orVariopol Cable to Model 5081-A

Strona 5 - CALIBRATION

FIGURE 13. Wiring Sensor with Standard Cable toModel 1066 Transmitter.FIGURE 14. Wiring Sensor with Optimum EMI/RFI orVariopol Cable to Model 1066 T

Strona 6 - MAINTENANCE

MODEL 499A CL-01 MAINTENANCEMAINTENANCECLEANING THE MEMBRANE.Keep the membrane clean and free from dirt and algae.Clean the membrane with water spraye

Strona 7 - SPARE PARTS

33523-00 Electrolyte Fill Plug 9550094 O-Ring, Viton 2-01433521-00 Membrane Retainer23501-08 Free Chlorine Membrane Assembly: includes one membrane as