Modbus
®
Mapping Assignments for Micro Motion
®
Transmitters 23
Modbus Mapping Assignments continued
Quaternary variable codes (holding register 40015)
0 Mass flow rate
ÖÖÖ
1 Temperature
ÖÖÖ
2 Mass total
ÖÖÖ
3 Density
ÖÖÖ
4 Mass inventory
ÖÖÖ
5 Volume flow rate
ÖÖÖ
6 Volume total
ÖÖÖ
7 Volume inventory
ÖÖÖ
15 API: Temperature-corrected density
ÖÖÖ
16 API: Temperature-corrected (standard) volume flow
ÖÖÖ
17 API: Temperature-corrected (standard) volume total
ÖÖÖ
18 API: Temperature-corrected (standard) volume inventory
ÖÖÖ
19 API: Batch-weighted average corrected density
ÖÖÖ
20 API: Batch-weighted average temperature
ÖÖÖ
33 API: CTL
ÖÖÖ
47 Drive gain
ÖÖÖ
53 Externally read pressure
ÖÖÖ
55 Externally read temperature
ÖÖÖ
Event alarm type codes (holding register 40139 or 40140)
1 High alarm
Ö ÖÖÖ
2 Low alarm
Ö ÖÖÖ
Flow direction codes (holding register 40017)
0 Forward flow only
Ö ÖÖÖ
1 Reverse flow only
Ö ÖÖÖ
2 Bidirectional flow
Ö ÖÖÖ
3 Absolute forward/reverse
ÖÖÖ
4 Negate - forward only
Ö
5 Negate - bidirectional
Ö
Process variable and diagnostic codes for slot addresses (holding registers 40751-40782)
0 Mass flow rate
ÖÖÖ
1 Temperature
ÖÖÖ
2 Mass totalizer
ÖÖÖ
3 Density
ÖÖÖ
4 Mass inventory
ÖÖÖ
5 Volume flow rate
ÖÖÖ
6 Volume totalizer
ÖÖÖ
7 Volume inventory
ÖÖÖ
10 Event 1
ÖÖÖ
11 Event 2
ÖÖÖ
12 Status word 1 (419/420)
ÖÖÖ
13 Status word 2 (421/422)
ÖÖÖ
14 Status word 3 (423/424)
ÖÖÖ
15 API: Temperature-corrected density
ÖÖÖ
16 API: Temperature-corrected (standard) volume flow
ÖÖÖ
17 API: Temperature-corrected (standard) volume total
ÖÖÖ
18 API: Temperature-corrected (standard) volume inventory
ÖÖÖ
19 API: Batch-weighted average corrected density
ÖÖÖ
20 API: Batch-weighted average temperature
ÖÖÖ
Table 2-8. Integer codes continued