Model Solu Comp
®
II
Dual Input pH/Conductivity Analyzer
Instruction Manual
PN 51-1055pHC/rev.K
February 2006
Przeglądanie stron 0
1 2 3 4 5 6 ... 86 87

Podsumowanie treści

Strona 1 - Model Solu Comp

Model Solu Comp®IIDual Input pH/Conductivity AnalyzerInstruction ManualPN 51-1055pHC/rev.KFebruary 2006

Strona 2 - WARNINGS

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONSSECTION 1.0. DESCRIPTION AND SPECIFICATIONS1.1 FEATURES AND APPLICATIONS1.2 SPECIFICAT

Strona 3 - QUICK START GUIDE

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS21.2 SPECIFICATIONS - GeneralCase: ABS (panel-mount), polycarbonate (pipe- andsurface-mou

Strona 4

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONSMeasures conductivity in the range 0 to 20,000 µS/cm.Display choices are conductivity, res

Strona 5 - QUICK REFERENCE GUIDE

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS4When used with Model Series 200 ToroidalConductivity Sensors, display choices are conduct

Strona 6 - MODEL SOLU COMP II

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONSFor use with any standard pH or ORP sensor and allUniloc sensors and junction boxes with b

Strona 7 - TABLE OF CONTENTS CONT’D

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONSWhen used with a chlorine specific membrane-coveredamperometric sensor, display choices ar

Strona 8 - LIST OF FIGURES

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS7When used with an oxygen permeable membrane-covered amperometric sensor, display choices

Strona 9 - About This Document

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS81.3 ORDERING INFORMATIONThe Solu Comp II analyzers offer the choice of single or dual sen

Strona 10 - SECTION 1.0

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS9COMBINATION GUIDE55* For D.O. sensors with 22k thermistor, use Suite 1 only.

Strona 11 - 1.2 SPECIFICATIONS - General

MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS10ACCESSORIES (Weights are rounded up to nearest whole lb or 0.5 kg)PART NUMBER DESCRIPTIO

Strona 12

ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PROCEEDING!Your purchase from Rosemount Analytical, Inc. hasresulted in one of the finest instruments ava

Strona 13 - S2 847.1µS/cm 100 C

11SECTION 2.0. INSTALLATIONMODEL SOLU COMP II SECTION 2.0INSTALLATION2.1 UNPACKING AND INSPECTION2.2 INSTALLATIONType of Mounting SectionPanel 2.2.2

Strona 14

12MODEL SOLU COMP II SECTION 2.0INSTALLATIONFIGURE 2-1. Panel Mount InstallationAccess to the wiring terminals is through the rear cover. Four screws

Strona 15

MODEL SOLU COMP II SECTION 2.0INSTALLATIONFIGURE 2-2. Pipe Mount InstallationThe front panel is hinged at the bottom. The panel swings down for access

Strona 16 - DISSOLVED OZONE (Code -26)

MODEL SOLU COMP II SECTION 2.0INSTALLATION14FIGURE 2-4. Surface Mount InstallationThe front panel is hinged at the bottom. The panel swings down for a

Strona 17 - 1.3 ORDERING INFORMATION

3.2 POWER, ALARM, OUTPUT, AND SENSOR CONNECTIONSThe Solu Comp II is available in two mounting configurations.The positions of the power, alarm, outpu

Strona 18 - COMBINATION GUIDE

16MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-2. Wiring Connections for Solu Comp II Model 1055-01-10 (Panel Mount with 1

Strona 19

MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-5. Wiring Connections for Solu Comp II Model 1055-02-11 (Surface/Pipe Mountin

Strona 20 - INSTALLATION

18MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-6. Panel Mount Wiring to Models 399-09-62, 400, 401, 402, 403, and 404.DWG. NO. REV.41055105 B

Strona 21 - MILLIMETER

MODEL SOLU COMP II SECTION 3.0WIRING19FIGURE 3-7. Pipe/Wall Mount Wiring to Models 399-14, 400, 401, 402, 403, and 404.DWG. NO. REV.41055106 B

Strona 22

20MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-8. Pipe/Wall Mount Wiring to Models 399VP-09, 400, 401, 402, 403, and 404.DWG. NO. REV.41055107 B

Strona 23

QUICK START GUIDEFOR MODEL SOLU COMP II pH/CONDUCTIVITY ANALYZER(Model Option 1055-22-30)1. Refer to Section 2.0 for installation instructions.2. Wi

Strona 24 - SECTION 3.0

MODEL SOLU COMP II SECTION 3.0WIRING21FIGURE 3-9. Panel Mount Wiring to Models 399VP-09, 400, 401, 402, 403, and 404.DWG. NO. REV.41055108 B

Strona 25

22MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-10. Pipe/Wall Mount Wiring to Models 399-09-62, 400, 401, 402, 403, and 404.DWG. NO. REV.41055109 B

Strona 26

MODEL SOLU COMP II SECTION 3.0WIRING23FIGURE 3-11. Panel Mount Wiring to Models 399-14, 400, 401, 402, 403, and 404.DWG. NO. REV.41055110 B

Strona 27 - DWG. NO. REV

24MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-12. Wiring Model 1055-11-22-32 to Models 399-09-62 and GP-1.DWG. NO. REV.40105536 C

Strona 28

MODEL SOLU COMP II SECTION 3.0WIRING25FIGURE 3-13. Wiring Model 1055-10-22-32 to Model 399-14DWG. NO. REV.40105559 EFIGURE 3-14. Wiring Model 1055 t

Strona 29

26MODEL SOLU COMP II SECTION 4.0DISPLAY AND OPERATIONSECTION 4.0DISPLAY AND OPERATION4.1 DISPLAY4.2 KEYPAD4.3 PROGRAMMING AND CALIBRATING THE SOLU

Strona 30

27MODEL SOLU COMP II SECTION 4.0DISPLAY AND OPERATION4.3 PROGRAMMING AND CALIBRATING THE SOLU COMP II- TUTORIALSetting up and calibrating the Solu Co

Strona 31

MODEL SOLU COMP II SECTION 4.0DISPLAY AND OPERATION1. If a security code has been programmed, pressing MENU causes thesecurity screen to appear.2. Ent

Strona 32

29MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZERSECTION 5.0PROGRAMMING THE ANALYZER5.1 GENERAL5.2 CHANGING STARTUP SETTINGS5.3 CONFIGURING

Strona 33

30MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZERTABLE 5-1. DEFAULT SETTINGS1. SENSOR-OUTPUT ASSIGNMENTS (pH, ORP, redox, conductivity, resist

Strona 34

5. Choose the desired language. Select >> to show more choices.6. Choose the number of sensors wired to the analyzer. NOTEIf One sensor is chose

Strona 35 - DISPLAY AND OPERATION

MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZERTABLE 5-1. DEFAULT SETTINGS (continued)6. MISCELLANEOUS SETTINGSSectionLanguage English 5.11Hol

Strona 36 - - TUTORIAL

32MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.3 CONFIGURING AND RANGING THE OUTPUTS.5.3.1 Purpose The Solu Comp II accepts input from a pH

Strona 37 - 4.5 USING HOLD

MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.3.3. Procedure: Configure Outputs.To choose a menu item, move the cursor to the item and press

Strona 38 - PROGRAMMING THE ANALYZER

34Alarm relays are single pole-double throw (SPDT). When an alarm is activated, the coil is energized. When an alarm activates, AL1, AL2, or AL3 (as a

Strona 39 - TABLE 5-1. DEFAULT SETTINGS

35MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.4.3 Procedure: Configuring AlarmsTo choose a menu item, move the cursor to the item and pres

Strona 40

36MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.4.4 Procedure: Programming Alarm SetpointsTo choose a menu item, move the cursor to the item

Strona 41

37MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.5 SELECTING THE TYPE OF MEASUREMENT AND ACTIVATING SOLUTION TEMPERATURE CORRECTIONS5.5.1 Pu

Strona 42

MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.5.3 Definitions — conductivity/resistivity1. NEUTRAL SALT CORRECTION. Neutral salt temperature

Strona 43

MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.5.4 Procedure.To choose a menu item, move the cursor to the item and press ENTER. To store a n

Strona 44

40MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.6 CHOOSING TEMPERATURE UNITS AND MANUAL OR AUTOMATIC TEMPERATURECOMPENSATION5.6.1 PurposeTh

Strona 45

QUICK REFERENCE GUIDEMENU TREE FOR pH/ORP/REDOX/CONDUCTIVITY/RESISTIVITY/TDS MEASUREMENTSpH/cond/temp

Strona 46 - TEMPERATURE CORRECTIONS

41MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.7 SETTING A SECURITY CODE5.7.1 Purpose.This section describes how to set a security code. Th

Strona 47

MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.8 NOISE REJECTION5.8.1 Purpose.For maximum noise rejection, the frequency of the ac power must

Strona 48

43MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.11 SELECTING A DEFAULT SCREEN, LANGUAGE, AND SCREEN CONTRAST5.11.1 PurposeThis section descr

Strona 49 - COMPENSATION

44MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.11.4 Procedure: Changing Screen Contrast1. Press MENU. The main menu screen appears. Choose

Strona 50 - 5.7 SETTING A SECURITY CODE

45MODEL SOLU COMP II SECTION 6.0CALIBRATIONSECTION 6.0CALIBRATION6.1 INTRODUCTION6.2 CALIBRATING TEMPERATURE6.3 AUTO CALIBRATION - pH6.4 MANUAL

Strona 51 - 5.8 NOISE REJECTION

46MODEL SOLU COMP II SECTION 6.0CALIBRATION6.2 CALIBRATING TEMPERATURE6.2.1 PurposeTemperature is important in the measurement of pH, ORP, and conduct

Strona 52

476.2.2 Procedure1. Remove the sensor from the process. Place it in an insulated container of water along with a calibrated ther-mometer. Submerge at

Strona 53

MODEL SOLU COMP II SECTION 6.0CALIBRATION6.3 AUTO CALIBRATION — pH6.3.1 Purpose1. New sensors must be calibrated before use. Regular recalibration is

Strona 54 - CALIBRATION

49MODEL SOLU COMP II SECTION 6.0CALIBRATIONa. Press MENU. The main menu screen appears. Choose Calibrate.b. Choose Sensor1 (pH sensor). c. Choose Meas

Strona 55 - 6.2 CALIBRATING TEMPERATURE

50MODEL SOLU COMP II SECTION 6.0CALIBRATION6.4 MANUAL CALIBRATION — pH6.4.1 Purpose1. New sensors must be calibrated before use. Regular recalibration

Strona 56

iMODEL SOLU COMP II TABLE OF CONTENTSMODEL SOLU COMP IIDUAL INPUT pH/CONDUCTIVITY ANALYZERTABLE OF CONTENTSSection Title Page1.0 DESCRIPTION AND SPECI

Strona 57 - 6.3 AUTO CALIBRATION — pH

51MODEL SOLU COMP II SECTION 6.0CALIBRATIONa. Press MENU. The main menu screen appears. Choose Calibrate.b. Choose Sensor1 (pH sensor). c. Choose Meas

Strona 58

52MODEL SOLU COMP II SECTION 6.0CALIBRATION6.5 STANDARDIZATION — pH6.5.1 Purpose1. The pH measured by the Solu Comp II analyzer can be changed to mat

Strona 59 - 6.4 MANUAL CALIBRATION — pH

MODEL SOLU COMP II SECTION 6.0CALIBRATION6.6 ENTERING A KNOWN SLOPE VALUE — pH6.6.1 PurposeIf the electrode slope is known from other measurements, it

Strona 60

54MODEL SOLU COMP II SECTION 6.0CALIBRATION6.7 ORP CALIBRATION6.7.1 Purpose1. For process control, it is often important to make the measured ORP agre

Strona 61 - 6.5 STANDARDIZATION — pH

6.8 CALIBRATING AN IN-SERVICE SENSOR — CONDUCTIVITY6.8.1 Purpose1. After a conductivity sensor has been in service for aperiod of time, recalibration

Strona 62

56MODEL SOLU COMP II SECTION 6.0CALIBRATIONa. Press MENU. The main menu screen appears. Choose Program.b. Choose Measurement.c. Choose Sensor2 (conduc

Strona 63 - 6.7 ORP CALIBRATION

MODEL SOLU COMP II SECTION 6.0CALIBRATIONPossible Cal ErrorProceed?YYeessNoUpdated S2 CellConst: 1.0013/cmh. If the change is more than 10% of the pre

Strona 64 - To calibrate a 0.01/cm sensor

58MODEL SOLU COMP II SECTION 6.0CALIBRATIONa. Press MENU. The main menu screen appears. Choose Program.b. Choose Measurement.c. Choose Sensor2 (conduc

Strona 65

MODEL SOLU COMP II SECTION 6.0CALIBRATIONCalibrate?SSeennssoorr11Sensor2Cal Sensor2?MMeeaassuurreemmeennttTempCalS2? InProcessMeterCCeellllCCoonnssttC

Strona 66

60MODEL SOLU COMP II SECTION 6.0CALIBRATION6.10 CALIBRATING THE ANALYZER - CONDUCTIVITY.6.10.1 Purpose1. After a conductivity analyzer has been in ser

Strona 67

iiSection Title Page6.0 CALIBRATION ... 456.1 Int

Strona 68

MODEL SOLU COMP II SECTION 6.0CALIBRATION9. If a resistance value outside the range 100 Ω to 180 kΩ was entered, thecalibration will not be accepted.

Strona 69

62MODEL SOLU COMP II SECTION 7.0MAINTENANCESECTION 7.0MAINTENANCE7.1 OVERVIEW7.2 REPLACEMENT PARTS7.1 OVERVIEWThe Solu Comp II analyzer needs little

Strona 70 - SEE SECTION 5.10

MODEL SOLU COMP II SECTION 7.0MAINTENANCE63Location in ShippingFigure 7-1 PN Description Weight1 23823-00 Panel mounting kit, includes four brackets a

Strona 71 - MAINTENANCE

64MODEL SOLU COMP II SECTION 7.0MAINTENANCELocation in ShippingFigure 7-2 PN Description Weight1 note Screw, 6-32 x 1.38 in.2 note Washer3 23834-00 En

Strona 72

65MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTINGSECTION 8.0TROUBLESHOOTING8.1 OVERVIEW8.2 TROUBLESHOOTING USING FAULT CODES8.3 TROUBLESHOOTING WHEN N

Strona 73

MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.2.2 pH Sensitive Glass Membrane is BrokenThe Solu Comp II continuously measures the impedance between

Strona 74 - TROUBLESHOOTING

MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.3 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — pH.Problem See SectionNew temperature during cali

Strona 75

MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.3.3 Calibration Error during Standardization. During standardization, the millivolt signal from the pH

Strona 76 - Glass Imp 123M^

69MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.3.7 Calibration Was Successful, but Process pH Is Grossly Wrong and/or Noisy.Grossly wrong or noisy r

Strona 77

MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.4 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — CONDUCTIVITY8.4.1 Calibration Error Using Standa

Strona 78

MODEL SOLU COMP II TABLE OF CONTENTSLIST OF FIGURESNumber Title Page2-1 Panel Mount Installation ...

Strona 79

MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.4.5 Measurement seems wrong.A. If sensor is being installed for the first time, check the wiring conne

Strona 80

72MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.6 SIMULATING INPUTS — pH8.6.1 GeneralThis section describes how to simulate a pH input into the Solu

Strona 81 - 8.6 SIMULATING INPUTS — pH

MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.7 SIMULATING INPUTS — CONDUCTIVITYTo check the performance of the analyzer, use standard resistors to

Strona 82

Temp. (°C) Pt 100 (ΩΩ) Pt 1000 (ΩΩ)0 100.0 100010 103.9 103920 107.8 107825 109.7 109730 111.7 111740 115.5 115550 119.4 119460 123.2 123270 127.1 127

Strona 83 - 8.8 SIMULATING TEMPERATURE

MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTINGFIGURE 8-7. Checking for a Poisoned ReferenceElectrode.Refer to the sensor wiring diagram to identify the

Strona 84

76MODEL SOLU COMP II SECTION 9.0RETURN OF MATERIALSECTION 9.0RETURN OF MATERIAL9.1 GENERAL.To expedite the repair and return of instruments, proper co

Strona 85 - RETURN OF MATERIAL

WARRANTYSeller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufacturedor Services pro

Strona 86 - Irvine, CA 92606

Credit Cards for U.S. Purchases Only.The right people,the right answers,right now.ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITEhttp://www.raihome.com

Strona 87 - The right people

About This DocumentThis manual contains instructions for installation and operation of the Solu Comp IIModel 1055 Dual Input pH/Conductivity Analyzer.