Emerson Rosemount 2051 Wireless Pressure Transmitters Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Pomiary, testowanie i kontrola Emerson Rosemount 2051 Wireless Pressure Transmitters. Emerson Rosemount 2051 Wireless Pressure Transmitters User's Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Protocol

Reference Manual00809-0100-4102, Rev AAMay 2013Rosemount 2051 Wireless Pressure TransmittersPressure, Level, and Flow Solutions with WirelessHART™ Pro

Strona 2

ivReference Manual00809-0100-4102, Rev AATable of ContentsMay 2013Table of Contents AAppendix A: Specifications and Reference DataA.1 Performance Spec

Strona 3

88Reference Manual00809-0100-4102, Rev AASection 6: TroubleshootingMay 2013Troubleshooting6.3 Removing from serviceFollow these steps:1. Follow all pl

Strona 4

Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 201389Specifications and Reference DataAppendix A Specifications and Reference D

Strona 5

90Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataReference Accuracy(1)Total PerformanceMode

Strona 6

91Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataLong Term StabilityDynamic PerformanceLin

Strona 7 - Contents

92Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataAmbient Temperature Effect per 50°F (28°C)

Strona 8 - 3Section 3: Installation

93Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataA.2.2 Range and Sensor LimitsZero and Spa

Strona 9 - 6Section 6: Troubleshooting

94Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataUpdate rateUser selectable, 1 second to 60

Strona 10 - Reference Data

95Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataRosemount 2051LLimit is flange rating or

Strona 11 - Table of Contents

96Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataProcess Temperature LimitsAt atmospheric p

Strona 12

97Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataA.4 Physical specificationsA.4.1 Electric

Strona 13 - Section 1 Introduction

vReference Manual 00809-0100-4102, Rev AATable of ContentsMay 2013Table of Contents CAppendix C: Field Communicator Menu Trees and Fast KeysC.1 Field

Strona 14

98Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataProcess flanges and adaptersPlated carbon

Strona 15

99Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataBoltsPlated carbon steel per ASTM A449, T

Strona 16 - 1.4 Transmitter overview

100Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataA.4.6 Shipping Weights for 2051 Wireless

Strona 17 - 1.5.1 Wireless considerations

101Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataA.5 Dimensional DrawingsDimensions are i

Strona 18 - 1.5.3 Electrical

102Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference Data(1) Tolerances are 0.040 (1,02), –0.020

Strona 19 - 1.6 Service support

103Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataA.6 Ordering InformationTable A-6. 205

Strona 20

104Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataO-ringStandard StandardA Glass-filled PTF

Strona 21 - Section 2 Configuration

105Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataSeal Assemblies(9) Standard StandardS1(7

Strona 22 - Warnings ( )

106Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataCalibration CertificateStandard StandardQ

Strona 23 - 2.4 Basic setup

107Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataRC1/4 RC1/2 Process ConnectionExpandedD9

Strona 24 - 2.4.3 Configure update rate

viReference Manual00809-0100-4102, Rev AATable of ContentsMay 2013Table of Contents

Strona 25 - 2.5 Configure for Pressure

108Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataTable A-7. 2051T In-Line Pressure Trans

Strona 26 - 2.5.2 Set range points

109Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataMounting BracketStandard StandardB4 Brac

Strona 27 - 2, 2, 4, 2

110Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataSurface FinishStandard StandardQ16 Surfac

Strona 28

111Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference Data Table A-8. Rosemount 2051L Liqui

Strona 29 - DP Flow example

112Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataP 2-in. ANSI/ASME B16.5 Class 600 CS ★E 3

Strona 30

113Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataO-ringStandardStandardA Glass-filled PTF

Strona 31 - 2.6.3 Set range points

114Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataToolkit Total System Performance ReportsS

Strona 32 - 2.7 Review configuration data

115Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataTable A-9. Rosemount 2051CFA Annubar Fl

Strona 33 - Fast Keys

116Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataPipe Material / Mounting Assembly Materia

Strona 34

117Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataR6 600# RTJ FlangeR9(1)900# RTJ FlangeRF

Strona 35 - 2.9.2 Damping

1Reference Manual 00809-0100-4102, Rev AASection 1: IntroductionMay 2013IntroductionSection 1 IntroductionUsing this manual . . . . . . . . . . . . .

Strona 36 - 2.10 Diagnostics and service

118Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataOptions (Include with selected model numb

Strona 37 - 2.10.2 Join status

119Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataExpandedG1 Needle Valves, Carbon SteelG3

Strona 38 - Communicator

120Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataTable A-10. Rosemount 2051CFC Compact Fl

Strona 39 - Field Communicator

121Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataTransmitter Housing Material Conduit Ent

Strona 40 - AMS creating a Reusable Copy

122Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataTransmitter Calibration CertificationStan

Strona 41 - Section 3 Installation

123Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataTable A-11. Rosemount 2051CFP Integral

Strona 42 - 3.2.1 Warnings ( )

124Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference Data0250 0.250-in. (6.35 mm) for 1-in. Pipe ★

Strona 43 - 3.3 Considerations

125Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataStandardWP5 Internal Antenna, Compatible

Strona 44 - Side mounted

126Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataTransmitter Calibration CertificationStan

Strona 45 - Installation

127Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataA.7 OptionsStandard configurationUnless

Strona 46 - Reference side filtering

2Reference Manual00809-0100-4102, Rev AASection 1: IntroductionMay 2013Introduction1.2.3 Rosemount 2051L Level Transmitter Measures level and specifi

Strona 47 - 3.4 Installation procedures

128Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataOutput informationOutput range points mus

Strona 48 - 3.4.1 Mount the transmitter

129Reference Manual 00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataRosemount 2051C Traditional Flange brack

Strona 49 - Flange bolts

130Reference Manual00809-0100-4102, Rev AAAppendix A: Reference DataMay 2013Specifications and Reference DataA.8 Spare partsMeter Kit 02051-9020-0001M

Strona 50 - Bolt installation

Reference Manual 00809-0100-4102, Rev AAAppendix B: Product CertificationsMay 2013131Product CertificationsAppendix B Product CertificationsWireless C

Strona 51

132Reference Manual00809-0100-4102, Rev AAAppendix B: Product CertificationsMay 2013Product CertificationsB.1.5 Ordinary location certification for FM

Strona 52 - Section 3: Installation

133Reference Manual 00809-0100-4102, Rev AAAppendix B: Product CertificationsMay 2013Product CertificationsSpecial Conditions for Safe Use (X)The plas

Strona 53 - 3.4.2 Impulse piping

134Reference Manual00809-0100-4102, Rev AAAppendix B: Product CertificationsMay 2013Product Certifications

Strona 54 - Mounting Requirements

Reference Manual 00809-0100-4102, Rev AAAppendix C: Field Communicator Menu TreesMay 2013135Field Communicator Menu Trees and Fast KeysAppendix C Fiel

Strona 55 - 3.4.3 Process connections

136Reference Manual00809-0100-4102, Rev AAAppendix C: Field Communicator Menu TreesMay 2013Field Communicator Menu Trees and Fast KeysFigure C-2. Rose

Strona 56

137Reference Manual 00809-0100-4102, Rev AAAppendix C: Field Communicator Menu TreesMay 2013Field Communicator Menu Trees and Fast KeysFigure C-3. Ros

Strona 57 - Process Connection

3Reference Manual 00809-0100-4102, Rev AASection 1: IntroductionMay 2013Introduction1.3 WirelessHART installation flowchartFigure 1-1. WirelessHART in

Strona 58

138Reference Manual00809-0100-4102, Rev AAAppendix C: Field Communicator Menu TreesMay 2013Field Communicator Menu Trees and Fast KeysFigure C-4. Rose

Strona 59

Reference Manual 00809-0100-4102, Rev AAAppendix D: Network design best practicesMay 2013139Appendix D: Network design best practicesAppendix D Networ

Strona 60

140Reference Manual00809-0100-4102, Rev AAAppendix D: Network design best practicesMay 2013Appendix D: Network design best practicesFor examples and c

Strona 62 - 3.5.4 Manifold operation

Reference Manual00809-0100-4102, Rev AAMay 2013Standard Terms and Conditions of Sale can be found at www.rosemount.com/terms_of_saleThe Emerson logo i

Strona 63

4Reference Manual00809-0100-4102, Rev AASection 1: IntroductionMay 2013Introduction1.4 Transmitter overviewThe Rosemount 2051C Coplanar design is offe

Strona 64

5Reference Manual 00809-0100-4102, Rev AASection 1: IntroductionMay 2013IntroductionFigure 1-3. Block diagram of operation A. Sensor ModuleB. Electron

Strona 65

6Reference Manual00809-0100-4102, Rev AASection 1: IntroductionMay 2013IntroductionNetwork design best practicesWhen mounting the device, recommended

Strona 66

7Reference Manual 00809-0100-4102, Rev AASection 1: IntroductionMay 2013Introduction701PGNKF). Each battery contains approximately 5 grams of lithium.

Strona 68 - 4.2.1 Warnings ( )

8Reference Manual00809-0100-4102, Rev AASection 1: IntroductionMay 2013IntroductionThe Power Module contains a primary lithium-thionyl chloride batter

Strona 69 - 4.4 Verifying operation

Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 20139ConfigurationSection 2 ConfigurationOverview . . . . . . . . . . . . . . . .

Strona 70 - Smart Wireless Gateway

10Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013ConfigurationWarnings ( )2.3 Required bench top configurationBench top config

Strona 71 - Troubleshooting

11Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.3.1 Connection diagrams Bench hook-upConnect the bench equipm

Strona 72 - Network

12Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.4.2 Join device to networkIn order to communicate with the Sma

Strona 73 - HART lock

13Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.4.4 Set process variable unitsThe PV Unit command sets the pr

Strona 74 - 2, 2, 7, 2

14Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013ConfigurationTable 2-1. Variable MappingNoteThe variable assigned to the pri

Strona 75 - 5.3 Calibration overview

15Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.5.3 Set transmitter percent of range (transfer function)The R

Strona 76

16Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.6 Configure for Level and Flow2.6.1 Configuring scaled variabl

Strona 77 - Digital Zero

17Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configurationa. When configuring for level applications, select Linear under

Strona 78 - Sample calculation for 2051

iReference Manual 00809-0100-4102, Rev AARosemount 2051 Wireless Pressure, Flow, and Level SolutionsRead this manual before working with the product.

Strona 79 - 5.4 Trim the pressure signal

18Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configurationrecommended to use the low flow cutoff function in order to have

Strona 80 - 5.4.2 Sensor Trim

19Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration1. Select Manual Setup and click on the HART tab.2. Assign Prim

Strona 81

20Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.7 Review configuration dataThe following is a list of factory

Strona 82 - Span Effect

21Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.7.3 Review radio informationTo view radio information:1. From

Strona 83 - HT = URV + S x (URV) x P

22Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.8 Configuring the LCD displayThe LCD Display configuration com

Strona 84

23Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.9 Detailed transmitter setup2.9.1 Configure process alertsPro

Strona 85 - 5 8. 0

24Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2051 Wireless pressure transmitter, damping only takes effect wh

Strona 86 - 2 5. 2 5

25Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.10.1 Device reset The master reset function will reset the de

Strona 87 -

26Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.10.3 Number of available neighbors Viewing number of available

Strona 88

27Reference Manual 00809-0100-4102, Rev AASection 2: ConfigurationMay 2013Configuration2.11 Advanced Functions for HART Protocol2.11.1 Saving, Recalli

Strona 89 - S R C H N G

iiReference Manual 00809-0100-4102, Rev AAExplosions could result in death or serious injury: Installation of this transmitter in an explosive environ

Strona 90

28Reference Manual00809-0100-4102, Rev AASection 2: ConfigurationMay 2013ConfigurationWhen finished, the Field Communicator informs you of the status.

Strona 91 - P S I

Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 201329InstallationSection 3 InstallationOverview . . . . . . . . . . . . . . . . .

Strona 92 - 3. 6 0

30Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013Installation3.2.1 Warnings ( )Failure to follow these installation guidelines

Strona 93 - f a i l u r

31Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013Installation3.3 Considerations3.3.1 Installation considerationsMeasurement pe

Strona 94 - w a r n

32Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationInternal antenna positionThe internal antenna is designed for mult

Strona 95 - Section 6 Troubleshooting

33Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-2. Installation Flowchart3.3.5 Draft range consideration

Strona 96 - 6.2.1 Warnings ( )

34Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationTilting of the transmitter may cause a zero shift in the transmitt

Strona 97

35Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013Installation3.4 Installation proceduresFor dimensional drawing information re

Strona 98

36Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationEnvironmental seal for housingAlways ensure a proper seal by insta

Strona 99

37Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-5. Mounting bracket option code B4A. 5/16 x 11/2 Bolts f

Strona 100 - 6.3 Removing from service

iiiReference Manual 00809-0100-4102, Rev AAShipping considerations for wireless products (Lithium Batteries: Green Power Module, model number 701PGNKF

Strona 101

38Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationBolt installationOnly use bolts supplied with the Rosemount 2051 o

Strona 102 - Total Performance

39Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-7. Mounting bolts and bolt configurations for coplanar f

Strona 103 - Dynamic Performance

40Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-9. Panel mounting bracket option codes B2 and B8Figure 3-

Strona 104 - A.2 Functional Specifications

41Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013Installation3.4.2 Impulse pipingBest practicesThe piping between the process

Strona 105 - A.2.2 Range and Sensor Limits

42Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationMounting RequirementsRefer to Figure 3-11 on page 43 for examples

Strona 106 - A.3.1 Overpressure limits

43Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-11. Installation examplesA. Drain/vent valves3.4.3 Proce

Strona 107 - A.3.4 Temperature limits

44Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013Installation1. Remove the flange bolts.2. Leaving the flange in place, move th

Strona 108 - A.3.7 Damping

45Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-12. Inline gage low side pressure portA. Low side pressu

Strona 109 - A.4 Physical specifications

46Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationTo make connections, perform the following procedure:1. Remove the

Strona 110 - A.4.5 Non-Wetted Parts

47Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-14. Optional LCD Display3.5 Rosemount 304, 305 and 306 i

Strona 111 - Power module

ivReference Manual 00809-0100-4102, Rev AA

Strona 112 - May 2013

48Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-15. Integral Manifold DesignsA. 2051C and 304 Conventiona

Strona 113 - A.5 Dimensional Drawings

49Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFigure 3-16. Bolt tightening pattern3. If the PTFE sensor module

Strona 114

50Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationNoteIf using thread tape, be sure the thread tape does not strip w

Strona 115 - A.6 Ordering Information

51Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationDrain/VentValveDrain/Vent ValveEqualize(closed)ProcessIsolate(ope

Strona 116

52Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013InstallationFive-valve Natural Gas configurations shown:NoteDo not open the lo

Strona 117

53Reference Manual 00809-0100-4102, Rev AASection 3: InstallationMay 2013Installation3. Open the equalize valve on the low pressure (downstream) side

Strona 118

54Reference Manual00809-0100-4102, Rev AASection 3: InstallationMay 2013Installation

Strona 119

Reference Manual 00809-0100-4102, Rev AASection 4: CommissioningMay 201355CommissioningSection 4 CommissioningSafety messages . . . . . . . . . . . .

Strona 120 - 2–14 NPT Female ★

56Reference Manual00809-0100-4102, Rev AASection 4: CommissioningMay 2013Commissioning4.2.1 Warnings ( )Failure to follow these installation guideline

Strona 121

57Reference Manual 00809-0100-4102, Rev AASection 4: CommissioningMay 2013Commissioning4.3 Viewing network statusIf the Rosemount 2051 Wireless was co

Strona 122 - Standard Standard

iReference Manual 00809-0100-4102, Rev AATable of ContentsMay 2013Table of Contents 1Section 1: Introduction1.1 Using this manual. . . . . . . . . . .

Strona 123

58Reference Manual00809-0100-4102, Rev AASection 4: CommissioningMay 2013CommissioningFigure 4-1. Diagnostic Screen SequenceFigure 4-2. Network Join S

Strona 124

59Reference Manual 00809-0100-4102, Rev AASection 4: CommissioningMay 2013CommissioningFigure 4-3. Smart Wireless Gateway Explorer page. AMS Suite Wir

Strona 125

60Reference Manual00809-0100-4102, Rev AASection 4: CommissioningMay 2013CommissioningFigure 4-5. Smart Wireless Gateway Network Settings4.4.1 Using t

Strona 126

61Reference Manual 00809-0100-4102, Rev AASection 4: CommissioningMay 2013CommissioningTable 4-1. Rosemount 2051 Wireless Fast Key SequenceFigure 4-6

Strona 127

62Reference Manual00809-0100-4102, Rev AASection 4: CommissioningMay 2013CommissioningConfiguring HART Lock using Field CommunicatorConfiguring HART L

Strona 128

63Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceSection 5 Operation and maintenanceOve

Strona 129 - 2-in. NPT) ★

64Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceTable 5-1. Recommended Calibration Task

Strona 130

65Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceFor transmitters that are field install

Strona 131

66Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceSample calculation for 2051 (0.04% accur

Strona 132

67Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceExampleA range 4 differential pressure

Strona 133

iiReference Manual00809-0100-4102, Rev AATable of ContentsMay 2013Table of Contents2.6.2 Re-Mapping device variables . . . . . . . . . . . . . . . .

Strona 134

68Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceposition. Since this correction maintain

Strona 135

69Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceNoteUse a pressure input source that is

Strona 136

70Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenance5.4.3 Recall Factory Trim—Sensor TrimThe

Strona 137

71Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceZero Effect:± 0.1% of the upper range l

Strona 138 - Appendix A: Reference Data

72Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceTo calculate the low trim (LT) value: To

Strona 139 - A.7 Options

73Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenance5.5 LCD Screen Messages5.5.1 Startup Sc

Strona 140 - Configuration buttons

74Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceSV Screen - sensor temperature valueTV S

Strona 141

75Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenance5.5.2 Diagnostic Button Screen Sequence

Strona 142 - A.8 Spare parts

76Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenance5.5.3 Network Diagnostic Status ScreensT

Strona 143 - B.1 Wireless Certifications

77Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceDiagnostic Button Screen 4.4: the devic

Strona 144 - B.1.8 European certifications

iiiReference Manual 00809-0100-4102, Rev AATable of ContentsMay 2013Table of Contents3.5.1 Rosemount 305 Integral Manifold installation procedure . .

Strona 145

78Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceDiagnostic Button Screen 4.9: the device

Strona 146 - Product Certifications

79Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenance5.5.4 Device Diagnostic ScreensThe foll

Strona 147 - Trees and Fast Keys

80Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceQV Screen - voltage reading at the power

Strona 148

81Reference Manual 00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceDiagnostic Button Screen 7.1: the termi

Strona 149

82Reference Manual00809-0100-4102, Rev AASection 5: Operation and MaintenanceMay 2013Operation and maintenanceNoteUse the Rosemount Wireless LCD Part

Strona 150

Reference Manual 00809-0100-4102, Rev AASection 6: TroubleshootingMay 201383TroubleshootingSection 6 Troubleshooting Overview . . . . . . . . . . . .

Strona 151 - D.1 Effective range

84Reference Manual00809-0100-4102, Rev AASection 6: TroubleshootingMay 2013Troubleshooting6.2.1 Warnings ( )Table 6-1. Rosemount 2051 Wireless Device

Strona 152

85Reference Manual 00809-0100-4102, Rev AASection 5: TroubleshootingMay 2013TroubleshootingPressure has Exceeded LimitsThe sensor has exceeded the max

Strona 153

86Reference Manual00809-0100-4102, Rev AASection 6: TroubleshootingMay 2013TroubleshootingLO Alarm The primary variable has surpassed the user defined

Strona 154

87Reference Manual 00809-0100-4102, Rev AASection 5: TroubleshootingMay 2013TroubleshootingTable 6-2. Rosemount 2051 Wireless TroubleshootingTable 6-

Komentarze do niniejszej Instrukcji

Brak uwag