MODEL Xmt-T TOROIDAL CONDUCTIVITY TRANSMITTER
FEATURES
The Model Xmt-T measures conductivity or % concentration in the harshest environments. Transmitter can also be configured,
using the "Custom Curve" feature, to measure ppm, %, or a no unit variable according to a programmable conductivity vs.
variable curve. The transmitter recognizes the type of RTD (Pt100 or Pt1000) being used. Measurements are automatically
corrected for the resistance of the sensor cable to improve accuracy of high conductivity readings. The Xmt-T is prepro-
grammed with several concentration curves: 0-12% NaOH, 0-15% HCl, 0-20% NaCl, and 0-25% and 96-99.7% H
2
SO
4
. Tem-
perature compensation choices are linear slope correction or none (to provide display of raw conductivity).
FUNCTIONAL SPECIFICATIONS
Automatic Temperature Compensation:
3-wire Pt 100 RTD or Pt 1000 RTD
Conductivity: 0 to 200°C (32 to 392°F)
% Concentration: 0 to 100°C (32 to 212°F)
Diagnostics: The internal diagnostics can detect:
Calibration Error ROM Failure
Temperature Slope Error Zero Error
High Temperature Warning CPU Failure
Low Temperature Warning Input Warning
Once one of the above is diagnosed, the LCD will display a
message describing the problem.
Digital Communications:
HART: PV, SV, and TV assignable to measurement (con-
ductivity, resistivity, or concentration), temperature, and
raw conductivity. Raw conductivity is measured con-
ductivity before temperature correction.
Fieldbus: Three AI blocks assignable to measurement
(conductivity, resistivity, or concentration), temperature,
and raw conductivity. Raw conductivity is measured
conductivity before temperature correction. Execution
time 75 msec. One PID block; execution time 150
msec. Device type: 4087. Device revision: 1. Certified
to ITK 4.6.
TRANSMITTER SPECIFICATIONS @ 25°C
Measured Range: 50 to 2,000,000 µS/cm (see chart)
Repeatability: ± 0.25% of reading
Temperature Accuracy:
± 0.2°C between 0 and 50°C
± 0.5°C above 50°C
(excludes inaccuracies in sensor)
Temperature Slope Adjustment: 0-5%/° C
% Concentration Ranges:
Sodium Hydroxide: 0 to 12%
Hydrochloric Acid: 0 to 15%
Sulfuric Acid: 0 to 25% and 96.0 to 99.7%
Sodium Chloride: 0 to 20%
Ambient Temperature Coefficient:
± 0.1% of reading ±2µS/cm per °C
Maximum Cable Length: 100 ft (30 m)
LOOP SPECIFICATIONS
Loop Accuracy: With a standard Model 228 or 225 sensor
with 20' cable, following a single point calibration, laborato-
ry accuracy at 25°C can be as good as ±2% of reading
and ±50 µS/cm.
For optimum performance, standardize the sensor in the
process at the conductivity and temperature of interest.
Results under real process conditions, at different tem-
peratures, or using other sensors may differ from above.
Calibration: Calibrate against previously calibrated standard
sensor and analyzer, or calibrate against solution of known
conductivity.
RECOMMENDED SENSORS:
Model 222 Flow-Through
Model 225 Clean-In-Place (CIP)
Model 226 Submersion/Insertion
Model 228 Submersion/Insertion/Retractable
Model 242 Flow-Through
Model 245 Sanitary Flow-Through
Model 247 Submersion/Flow-Tee
HAZARDOUS LOCATION APPROVALS
Intrinsic Safety (with appropriate safety barrier):
Class I, II, III, Div. 1
Groups A-G
T4 Tamb = 50°C
Class I, II, III, Div. 1
Groups A-G
T4 Tamb = 50°C
ATEX 1180
II 1 G
Baseefa04ATEX0215X
EEx ia IIC T4
Tamb = 0°C to 50°C
Non-Incendive:
Class I, Div. 2, Groups A-D
Dust Ignition Proof
Class II & III, Div. 1, Groups E-G
NEMA 4/4X Enclosure
Class I, Div. 2, Groups A-D
Dust Ignition Proof
Class II & III, Div. 1, Groups E-G
NEMA 4/4X Enclosure
Values shown are for 25°C conductivity with a temperature slope of 2% per degree C. The maximum range value
will be lower for solutions with a higher temperature slope. Minimum conductivity depends on sensor.
RECOMMENDED RANGES FOR TOROIDAL SENSORS
Conductivity Sensor Model 226 228 225 222 (1in.) 222 (2 in.) 242 245 247
Nominal Cell Constant 1.0 3.0 3.0 6.0 4.0 * * 3.5
Minimum Conductivity (μS/cm) 50 200 200 500 500 100* 100* 500
Maximum Conductivity (μS/cm) 1,000,000 2,000,000 2,000,000 2,000,000 2,000,000 2,000,000* 2,000,000 2,000,000
* Model 242 values depend on sensor configuration and wiring.