95
Reference Manual
00809-0100-4001, Rev JA
Section 5: Troubleshooting
November 2012
Troubleshooting
4k snsr EEPROM
error-factory ON
Replace the transmitter.
4k snsr EEPROM
error-user ON
Use the Field Communicator to reset the temperature
units and the calibration type.
Add item for ALL device types or only for
this ONE device type.
Asks the user whether the hot key item being added
should be added for all device types or only for the type of
device that is connected.
Command Not Implemented The connected device does not support this function.
Communication Error The communicator and the device are not
communicating correctly. Check all connections between
the Field Communicator and the device and resend the
information.
Configuration memory not compatible
with connected device
The configuration stored in memory is incompatible with
the device to which a transfer has been requested.
CPU board not initialized ON The electronics board is not initialized.
Replace the electronics board.
CPU EEPROM write failure ON Message sent to electronics board from HART signal
failed. Replace the electronics board.
Device Busy The connected device is busy performing another task.
Device Disconnected The device failed to respond to a command. Check all
connections between the Field Communicator and the
device and resend the command.
Device write protected Device is in write-protect mode. Data can not be written.
Device write protected. Do you still want
to shut off?
Device is in write-protect mode. Press YES to turn the Field
Communicator off and lose the unsent data.
Display value of variable on hotkey
menu?
Asks whether the value of the variable should be displayed
adjacent to its label on the hotkey menu if the item being
added to the hotkey menu is a variable.
Download data from configuration
memory to device
Press the SEND softkey to transfer information from the
communicator memory to the device.
Exceed field width Indicates that the field width for the current arithmetic
variable exceeds the device-specified description edit
format.
Exceed precision Indicates that the precision for the current arithmetic
variable exceeds the device-specified description edit
format.
Ignore next 50 occurrences of status? Select YES to ignore the next 50 occurrences of device
status, or select no to display every occurrence.
Illegal character An invalid character for the variable type was entered.
Illegal date The day portion of the date is invalid.
Illegal month The month portion of the date is invalid.
Illegal year The year portion of the date is invalid.
Incompatible CPU board and module ON Upgrade the electronics board or the sensor module to
the current revision.
Incomplete exponent The exponent of a scientific notation floating point
variable is incomplete.
Incomplete field The value entered is not complete for the variable type.
Message Description
Komentarze do niniejszej Instrukcji