Reference Manual
00809-0100-4001, Rev JA
Appendix A: Reference Data
November 2012
Specifications and reference data
A.1.10 Power supply effect
Less than ±0.005% of calibrated span per volt.
A.1.11 Electromagnetic compatibility (EMC)
Meets all relevant requirements of EN 61326 and NAMUR NE-21.
A.1.12 Transient protection (option code T1)
Meets IEEE C62.41, Category Location B
6 kV crest (0.5 s - 100 kHz)
3 kV crest (8 × 20 microseconds)
6 kV crest (1.2 × 50 microseconds)