Damping
4-20 mA HART
Analog output response to a step input change is user-enterable
from 0.0 to 60 seconds for one time constant. This software
damping is in addition to sensor module response time.
F
OUNDATION fieldbus
Transducer block: User configurable
AI Block: User configurable
PROFIBUS PA
AI Block only: User configurable
Physical specifications
Material selection
Emerson provides a variety of Rosemount products with various
product options and configurations including materials of
construction that can be expected to perform well in a wide
range of applications. The Rosemount product information
presented is intended as a guide for the purchaser to make an
appropriate selection for the application. It is the purchaser’s
sole responsibility to make a careful analysis of all process
parameters (such as all chemical components, temperature,
pressure, flow rate, abrasives, contaminants, etc.), when
specifying product materials, options, and components for the
particular application. Emerson Process Management is not in a
position to evaluate or guarantee the compatibility of the
process fluid or other process parameters with the product
options, configuration, or materials of construction selected.
Electrical connections
1
/2–14 NPT, G
1
/2, and M20 × 1.5 conduit. The polymer housing
(housing code P) has no conduit entries. HART interface
connections fixed to terminal block for output code A and to
701P Power Module for Output Code X.
Process connections
Rosemount 3051C
1
/4–18 NPT on 2
1
/8-in. centers
1
/2–14 NPT on 2-, 2
1
/8-, or 2
1
/4-in. centers
Rosemount 3051L
High pressure side: 2-, 3-, or 4-in., ASME B 16.5 (ANSI) Class 150,
300 or 600 flange; 50, 80 or 100 mm, PN 40 or 10/16 flange
Low pressure side:
1
/4–18 NPT on flange
1
/2–14 NPT on adapter
Rosemount 3051T
1
/2–14 NPT female.
G
1
/2 A DIN 16288 Male (Range 1–4 only)
Autoclave type F-250-C (Pressure relieved
9
/16–18 gland thread;
1
/4 OD high pressure tube 60° cone; available in SST for Range 5
transmitters only).
Rosemount 3051CF
For 3051CFA, see 00813-0100-4485 Rosemount 485 Annubar
For 3051CFC, see 00813-0100-4485 Rosemount 405 Compact
Orifice Plate
For 3051CFP, see 00813-0100-4485 Rosemount 1195 Integral
Orifice
Process-wetted parts
Drain/vent valves
316 SST, Alloy C-276, or Alloy 400 material (Alloy 400 not
available with 3051L)
Process flanges and adapters
Plated carbon steel
SST: CF-8M (Cast 316 SST) per ASTM A743
Cast C-276: CW-12MW per ASTM A494
Cast Alloy 400: M-30C per ASTM A494
Wetted O-rings
Glass-filled PTFE or Graphite-filled PTFE
Process isolating diaphragms
Rosemount 3051L process wetted parts
Flanged process connection (transmitter high side)
Process diaphragms, including process gasket surface
316L SST, Alloy C-276, or Tantalum
Extension
CF-3M (Cast version of 316L SST, material per ASTM-A743),
or Alloy C-276. Fits schedule 40 and 80 pipe.
Mounting flange
Zinc-cobalt plated CS or SST
Reference process connection (transmitter low side)
Isolating diaphragms
316L SST or Alloy C-276
Reference flange and adapter
CF-8M (Cast version of 316 SST, material per ASTM-A743)
Non-wetted parts
Electronics housing
Low-copper aluminum or CF-8M (Cast version of 316 SST).
Enclosure Type 4X, IP 65, IP 66, IP 68
Housing Material Code P: PBT/PC with NEMA 4X and IP66/67/68
Coplanar sensor module housing
SST: CF-3M (Cast 316L SST)
Isolating diaphragm material
3051CD
3051CG
3051T
3051CA
316L SST (UNS S31603)
Alloy C-276 (UNS N10276)
Alloy 400 (UNS N04400)
Tantalum (UNS R05440)
Gold-plated Alloy 400
Gold-plated 316L SST