Product Data Sheet
00813-0100-4530, Rev AA
Catalog 2008 - 2009
Level-21
Rosemount 5300 Series
Product Certifications
SAFETY NOTE
A safety isolator such as a zener barrier is always needed for
intrinsic safety.
Probes covered with plastic and/or with plastic discs may generate
an ignition-capable level of electrostatic charge under certain
extreme conditions. Therefore, when the probe is used in a
potentially explosive atmosphere, appropriate measures must be
taken to prevent electrostatic discharge.
Factory Mutual (FM) Approval
Project ID: 3020497
E5 Explosion Proof for Class I, Div. 1,
Groups B, C and D;
Dust Ignition Proof for Class II/III, Div. 1, Groups E, F and G;
With Intrinsically Safe connections to
Class I, II, III, Div. 1, Groups B, C, D, E, F and G.
Temp. Code T4
Ambient temperature limits: -40°C to +70°C
(1)
.
Seal not required.
I5, IE Intrinsically Safe for Class I, II, III, Div. 1, Groups A, B, C, D,
E, F and G,
Class I, Zone 0, AEx ia IIC T4 when installed per Control
Drawing: 9240 030-936.
Non-Incendive Class I, Div. 2, Groups A, B, C and D;
Suitable for Class II, III, Div. 2, Groups F and G.
4-20 mA / HART
®
model: U
i
=30 V dc, I
i
=130 mA, P
i
=1.0 W,
C
i
=7.26 nF, L
i
=0 H.
F
OUNDATION™ fieldbus model: U
i
=30 V dc, I
i
=300 mA,
P
i
=1.3 W, C
i
=0 nF, L
i
=0 H.
FISCO model: U
i
=17.5 V dc, I
i
=380 mA, P
i
=5.32 W, L
i
=C
i
=0.
Max operation:
4-20 mA / HART
®
model: 42.4 V, 25 mA,
F
OUNDATION™ fieldbus model: 32 V, 25 mA.
Temp. Code T4
Ambient temperature limits: -50°C to +70°C
(1)
ATEX Approval
Nemko 04ATEX1073X
SPECIAL CONDITIONS FOR SAFE USE (X)
The intrinsically safe circuits do not withstand the 500 V ac test as
specified in EN 50020 clause 6.4.12.
Probes covered with plastic and/or with plastic discs will have a
non-conducting area that exceeds the maximum permissible areas
for Group IIC and Category II 1G according to EN 50284 clause
4.4.3 (4 cm
2
). Therefore, when the antenna is used in a potentially
explosive atmosphere, appropriate measures must be taken to
prevent electrostatic discharge.
Impact and friction hazards need to be considered according to
EN 50284, clause 4.3.1 when the transmitter exposed to the
exterior atmosphere of the tank is made of light metal alloys, and is
used in Category II 1 G.
E1 Flame Proof:
II 1/2 GD T73°C
(2)
.
EEx iad IIC T4 (-40°C<T
a
<+70°C
(1)
)
U
m = 250 V.
I1, IA
Intrinsically Safe:
II 1 GD T73°C
(2)
.
EEx ia IIC T4 (-50°C<T
a
<+70°C
(1)
).
4-20 mA / HART
®
model: U
i
=30 V dc, I
i
=130 mA, P
i
=1.0 W,
C
i
=7.26 nF, L
i
=0 H.
F
OUNDATION™ fieldbus model: U
i
=30 V dc, I
i
=300 mA,
P
i
=1.5 W, C
i
=0 nF, L
i
=0 H.
FISCO model: U
i
=17.5 V dc, I
i
=380 mA, P
i
=5.32 W, L
i
=C
i
=0.
Installation Drawing: 9240 030-938
(1) +60°C with FOUNDATION™ fieldbus or FISCO option. (2) +63°C with FOUNDATION™ fieldbus or FISCO option.