Product Data Sheet
00813-0100-4530, Rev AA
Catalog 2008 - 2009
Level-9
Rosemount 5300 Series
Transition Zones
Transition zones are areas where measurements are
non-linear or will have reduced accuracy. See picture
and table below.
If measurements are desired at the very top of the
tank it is possible to mechanically extend the nozzle
and use the coaxial probe. Then the upper transition
zone is moved into the extension.
NOTE
The 4-20 mA set points are recommended to be
configured between the transition zones, within the
measuring range (see picture and diagram above).
Coaxial Rigid Twin Lead Flexible Twin Lead Rigid Single Lead Flex Single Lead
SST Probe XXXXX
Hastelloy Probe XX
Monel Probe XX
PTFE Covered Probe XX
(1)
HTHP Probe (SST) XXX
(1)
HP Probe (SST) XXX
(1)
(1) For measurements on liquids only. Consult factory if option is needed for solids.
For a flexible single lead probe with chuck, the lower
transition zone is measured upwards from the upper
clamp.
Upper Reference Point
Upper Transition Zone
Maximum
Measuring Range
Lower Transition Zone
Lower Reference Point
Dielectric Constant Coaxial Rigid Twin Lead Flexible Twin Lead Rigid Single Lead Flexible Single Lead
Upper
(1)
Transition
Zone
80 4.3 in. (11 cm) 4.3 in. (11 cm) 4.7 in. (12 cm) 4.3 in. (11 cm) 4.3 in. (11 cm)
2 4.3 in. (11 cm) 5.5 in. (14 cm) 5.5 in. (14 cm) 6.3 in. (16 cm) 7.1 in. (18 cm)
Lower
(2)
Transition
Zone
80 0.4 in. (1 cm) 1.2 in. (3 cm) 2 in.
(3)
(5 cm
(3)
) 2 in. (5 cm) 0 in.
(3)(4)
(0 cm
(3)(4)
)
2 2 in. (5 cm) 4 in. (10 cm) 5.5 in.
(3)
(14 cm
(3)
) 2.8 in.
(5)
(7 cm
(5)
)2 in.
(3)
(5 cm
(3)
)
(1) The distance from the upper reference point where measurements have reduced accuracy, see picture above.
(2) The distance from the lower reference point where measurements have reduced accuracy, see picture above.
(3) Note that the weight length adds to non-measurable area and is not included in the diagram. See “Dimensional Drawings”.
(4) The measuring range for the PTFE covered Flexible Single Lead probe includes the weight when measuring on a high dielectric media.
(5) If using a stainless steel centering disc, the lower transition zone is 8 in. (20 cm). If using a PTFE centering disc, the lower transition zone is not affected.