Rosemount Process Radar
Dynamic Vapor Compensation (DVC)
For high pressure saturated steam applications with a varying pressure
and/or temperature, certain models of the Rosemount 5300 Series have a
built-in function (DVC) that automatically compensates for varying vapor
dielectric constants. Option "4V" or "4U" in model code.
It is important that the vapor compensation function is calibrated after
installation.
1. Install unit in pipe/chamber
2. The surface level must be at least 19.7 in. (0.5 m) below the end of the
reflector (It is recommended to empty the pipe/chamber and to ensure it
is free from vapor or condensate remains)
3. Perform the calibration at ambient pressure and temperature conditions,
i.e. when the vapor dielectric constant is close to 1.0
4. Click Calibrate Vapor Compensation, and follow instructions
For more information and instructions, see the Rosemount 5300 Series
Reference Manual (Document No. 00809-0100-4530) section “Dynamic
Vapor Compensation”.
Step 4. Verification Viewing Measurement Data
In Rosemount Radar Master
Rosemount Radar Master > Tools > Device Display