Reference Manual
00809-0100-4801, Rev FA
October 2010
2-19
Rosemount 3051S Series
Figure 2-10. Remote Mount
Display wiring diagram
NOTE
Wire colors provided above are per Madison AWM Style 2549 cable. Wire
color may vary depending on cable selected.
Madison AWM Style 2549 cable includes a ground shield. This shield must be
connected to earth ground at either the SuperModule or the Remote Display,
but not both.
Junction Box Housing Remote Mount Display
4-20 mA
(white) 24 AWG
(blue) 24 AWG
(black) 22 AWG
(red) 22 AWG