
Reference Manual
00809-0100-4801, Rev FA
October 2010
4-3
Rosemount 3051S Series
Calibrate the sensor
• Sensor Trim (page 4-6)
• Zero Trim (page 4-6)
Calibrate the 4–20 mA output
• 4–20 mA Output Trim (page 4-8); or
• 4–20 mA Output Trim Using Other Scale (page 4-9)
Figure 4-1 on page 4-3 illustrates 3051S transmitter data flow. Data flow can
be summarized in four major steps:
1. A change in pressure is measured by a change in the sensor output
(Sensor Signal).
2. The sensor signal is converted to a digital format that is understood by
the microprocessor (Analog-to-Digital Signal Conversion).
3. Corrections are performed in the microprocessor to obtain a digital
representation of the process input (Digital PV).
4. The Digital PV is converted to an analog value (Digital-to-Analog Signal
Conversion).
Figure 4-1 also identifies the approximate transmitter location for each
calibration task. Data flows from left to right, and a parameter change affects
all values to the right of the changed parameter.
Not all calibration procedures should be performed for each 3051S
transmitter. Some procedures are appropriate for bench calibration, but
should not be performed during field calibration. Table 4-1 identifies the
recommended calibration procedures for each type of 3051S transmitter for
bench or field calibration.
Figure 4-1. Transmitter Data
Flow with Calibration Options
Transmitter Ranged 0 to 100 inH
2
O
(0 to 0,25 bar)
A/D
(STEP 2)
MICRO
(STEP 3)
D/A
(STEP 4)
Output:
20.00 mA
Output: 100 in. H
2
O
Pressure
Source
SENSOR
(STEP 1)
Komentarze do niniejszej Instrukcji