Reference Manual
00809-0100-4801, Rev FA
October 2010
3-27
Rosemount 3051S Series
5. Check the reference meter installed in the test loop to verify that it
displays the commanded output value.
a. If the values match, the transmitter and the loop are configured and
functioning properly.
b. If the values do not match, the current meter may be attached to the
wrong loop, there may be a fault in the wiring, the transmitter may
require an output trim, or the reference meter may be
malfunctioning.
After completing the test procedure, the display returns to the loop test screen
to choose another output value or to end loop testing.
6. Select End and click Next to end loop testing.
7. Select Next to acknowledge the loop can be returned to automatic
control.
8. Select Finish to acknowledge the method is complete.
ADVANCED FUNCTIONS
FOR HART PROTOCOL
Saving, Recalling, and
Cloning Configuration
Data
Use the cloning feature of the Field Communicator or the AMS “User
Configuration” feature to configure several 3051S transmitters similarly.
Cloning involves configuring a transmitter, saving the configuration data, then
sending a copy of the data to a separate transmitter. Several possible
procedures exist when saving, recalling, and cloning configuration data. For
complete instructions refer to the Field Communicator manual (publication no.
00809-0100-4276) or AMS on-line guides. One common method is as follows:
Field Communicator v3.3
1. Confirm and apply configuration changes to the first transmitter.
NOTE
If transmitter configuration has not been modified, “SAVE” option in step 2 will
be disabled
2. Save the configuration data:
a. Select “SAVE” from the bottom of the Field Communicator screen.
b. Choose to save your configuration in either the “Internal Flash”
(default) or the “Configuration EM” (Configuration Expansion
Module).
c. Enter the name for this configuration file. The default name is the
transmitter tag number.
d. Select “SAVE”.
3. Power the receiving transmitter and connect with Field Communicator.
4. Access the HART Application menu by pressing the LEFT ARROW from
the HOME/ONLINE screen.
Traditional Fast Keys
left arrow,
1, 2
Device Dashboard
Fast Keys
3, 4, 3