28
Reference Manual
00809-0100-4803, Rev DA
Section 4: Operation and Maintenance
April 2013
Operation and Maintenance
Note
If the installation uses a manifold, see “Manifold Operation to Zero the Differential Pressure
Sensor” on page 20.
1. Inspect the SuperModule PTFE o-rings. If the o-rings are undamaged, they may be
reused. Emerson Process Management recommends reusing o-rings if possible. If the
o-rings show any signs of damage, such as nicks or cuts, they should be replaced (part
number 03151-9042-0001 for glass-filled PTFE and part number 03151-9042-0002 for
graphite-filled PTFE).
Note
If replacing the o-rings, be careful not to scratch or deface the o-ring grooves or the surface of
the isolating diaphragm when removing the damaged o-rings.
2. Install the process flange on the SuperModule process connection. To hold the process
flange in place, install the two alignment cap screws finger tight (these screws are not
pressure retaining). Do not overtighten; this will affect module-to-flange alignment.
3. Install the appropriate flange bolts.
a. If the installation requires a
1
/4–18 NPT connection(s), use four 1.75-in. flange bolts.
Finger tighten the bolts. Go to Step d.
b. If the installation requires a
1
/2–14 NPT connection(s), use flange adapters and four
2.88-in. process flange/adapter bolts.
c. Hold the flange adapters and adapter o-rings in place while finger-tightening the bolts.
d. Tighten the bolts to the initial torque value using a crossed pattern. See Tab le 4 -1 for
appropriate torque values.
e. Tighten the bolts to the final torque value using a crossed pattern. See Tab le 4- 1 for
appropriate torque values. When fully tightened, the bolts should extend through the
top of the module housing.
f. Torque alignment screws to 30 in-lbs. (3.4 N-m). If the installation uses a conventional
manifold, then install flange adapters on the process end of the manifold using the
1.75-in. flange bolts supplied with the transmitter.
Table 4-1. Bolt Installation torque values
4. If the SuperModule PTFE o-rings are replaced, re-torque the flange bolts and alignment
cap screws after installation to compensate for seating of the PTFE o-ring.
Bolt Material Initial Torque Value Final Torque Value
CS-ASTM-A449 Standard 300 in-lb. (34 N-m) 650 in-lb. (73 N-m)
316 SST—Option L4 150 in-lb. (17 N-m) 300 in-lb. (34 N-m)
ASTM-A-193-B7M—Option L5 300 in-lb. (34 N-m) 650 in-lb. (73 N-m)
Alloy K-500—Option L6 300 in-lb. (34 N-m) 650 in-lb. (73 N-m)
ASTM-A-453-660—Option L7 150 in-lb. (17 N-m) 300 in-lb. (34 N-m)
ASTM-A-193-B8M—Option L8 150 in-lb. (17 N-m) 300 in-lb. (34 N-m)