29
Reference Manual
00809-0100-4803, Rev DA
Section 3: Configuration
April 2013
Configuration
10. Click Next to continue with the flow configuration on the Primary Element Selection tab.
11. Proceed with the steps in “Primary element selection” on page 21.
Figure 3-14. Custom gas fluid properties tab
Custom liquid [Density (T)]
The Custom Liquid option should be used for fluids not in the database such as proprietary
fluids.
1. Expand the Liquid category.
2. Expand the Custom Liquid category.
3. Select the Custom Liquid [Density (T)] option.
4. Enter the Nominal and Operating Temperature Range. Engineering Assistant will use this
range to identify the temperature values at which the fluid properties are required.
5. Click Next to continue the flow configuration on the Fluid Properties tab.
6. Enter the Custom Liquid Density at the given temperatures. Note that all temperature
values may be edited except the minimum and maximum temperatures.
7. Enter the Reference Density at the reference temperature.