Reference Manual
00809-0100-4803, Rev DA
Section 5: Troubleshooting
April 2013
Troubleshooting
(LCD is blank) LCD Update Error The LCD is no longer powered. Examine the LCD Connector and reset the
LCD.
If the problem persists, first replace the LCD
then replace the feature board electronics if
necessary.
PT LIMIT Process Temperature Out of Limits The Process Temperature sensor is
exceeding the user defined sensor
limits.
Verify the process conditions and adjust
limits if necessary. Check the Process
Temperature sensor and replace if
necessary.
RVRSE FLOW Reverse Flow Detected The transmitter is measuring a
negative differential pressure.
Verify the process conditions and the
transmitter installation.
SNSR COMM ERROR Module Communication Failure Communication between the sensor
module and the feature board
electronics have been lost.
Verify the connection between the sensor
module and the feature board electronics.
Replace the SuperModule assembly and/or
feature board electronics if necessary.
SNSR INCOMP ERROR Sensor Module Incompatibility The SuperModule assembly is not
compatible with the feature board
electronics. The SuperModule
assembly is not equipped with a
differential pressure sensor or it is an
older revision of the sensor module.
Replace the SuperModule assembly with
one that is compatible with the 3051S
MultiVariable Transmitter Plantweb
Housing.
SNSR MISSING ERROR Sensor Missing The sensor mapped to the primary
variable is not present.
Remap the primary variable to a sensor that
is present.
SNSRT LIMIT Sensor Temperature Out of Limits The Sensor Module Temperature is
exceeding the sensor limits.
Verify ambient conditions are within the
sensor limits.
XMTR Info Non-Volatile Memory Warning Transmitter information data is
incomplete. Transmitter operation
will not be affected.
Replace the feature board electronics at
next maintenance shutdown.
XMTR Info Error Non-Volatile Memory Error Non-volatile data of the device is
corrupted.
Replace the feature board electronics.
(Other message)
(1)
Maintenance Required The transmitter may not be operating
properly and requires attention.
Check other warning messages.
(Other message)
(1)
mA Output Fixed The 4-20 mA Analog Output Signal is
fixed at a constant value and is not
representative of the HART primary
variable
Disable Loop Current Mode.
(Other message)
(1)
Primary variable out of limits The primary variable is outside the
range of the transmitter.
View other diagnostic messages to
determine which variable is out of limits
(Other message)
(1)
Non-primary variable out of limits A variable other than the primary
variable is outside the range of the
transmitter.
View other diagnostic messages to
determine which variable is out of limits
(LCD is reading normally) Configuration changed A modification has been made to the
device configuration using a host
other than AMS.
No action is required; message will clear
after a change is made using AMS.
(LCD is reading normally) Cold start Transmitter was restarted. No action is required; message will clear
automatically.
(1) LCD messages will vary as it is specific to the possible problem.
LCD Messages Host Diagnostic Message Possible Problems Recommended Actions