Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 101
Optional Configuration continued
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
8.16 Configuring device settings
The device settings are used to describe the flowmeter components. Table 8-13 lists
and defines the device settings.
Note: The HART device ID, which is displayed in some menus, can be set only once,
and is usually set at the factory to the device serial number. If the HART device ID
has not been set, its value is 0.
If you are entering a date:
With ProLink II, use the left and right arrows at the top of the calendar to select the year and
month, then click on a date
With a Communicator, enter a value in the form mm/dd/yyyy
Standard volume flow (API) ✓✓✓✓
Standard density (API) ✓✓
Standard volume total (API)
Standard volume inventory (API)
Average corrected density (API) ✓✓
CTL (API)
(1) API process variables are available only if the petroleum measurement application has been installed on the transmitter.
If the enhanced density application has been installed, enhanced density process variables will be available.
Table 8-13 Device settings
Parameter Description
HART tag
(1)
(1) Devices using HART protocol to communicate with the transmitter may use either the HART address (see Section 8.15.2) or the
HART tag. You may configure either or both, as required by your other HART devices.
Also called the “software tag.” Used by other devices on the network to identify and communicate with
this transmitter via HART protocol. The HART tag must be unique on the network. If the transmitter
will not be accessed using HART protocol, the HART tag is not required.
Maximum length: 8 characters.
Descriptor Any user-supplied description. Not used in transmitter processing, and not required.
Maximum length: 16 characters.
Message Any user-supplied message. Not used in transmitter processing, and not required.
Maximum length: 32 characters.
Date Any user-selected date. Not used in transmitter processing, and not required.
Table 8-12 Process variable assignments for PV, SV, TV, and QV (continued)
Series 1000 Series 2000
Process variable
(1)
PV SV TV QV PV SV TV QV
Model 1500 AN
Model 1700 AN
Model 1700 IS
Model 2500 CIO
Model 2700 AN
Model 2700 IS
Model 2700 CIO
Przeglądanie stron 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 263 264

Komentarze do niniejszej Instrukcji

Brak uwag