Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 87
Optional Configuration continued
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
Note: For gas applications, Micro Motion recommends a flow damping value of 3.2.
Before setting the damping values, review Sections 8.7.1 through 8.7.3 for information on how the
damping values interact with other transmitter measurements and parameters.
8.7.1 Damping and volume measurement
When configuring damping values, be aware that volume measurement is derived from mass and
density measurements; therefore, any damping applied to mass flow and density will affect volume
measurements. Be sure to set damping values accordingly.
8.7.2 Interaction with the added damping parameter
Both the primary mA output and the secondary mA output (if it is available on your transmitter) have
a damping parameter (added damping). If damping is configured for flow, density, or temperature, the
same process variable is assigned to an mA output, and added damping is also configured for the mA
output, the effect of damping the process variable is calculated first, and the added damping
calculation is applied to the result of that calculation.
See Section 6.5.5 for more information on the added damping parameter.
8.7.3 Interaction with the update rate
Flow and density damping values depend on the configured update rate (see Section 8.8). If you
change the update rate, the damping values are automatically adjusted.
Note: The specific process variable selected for the 100 Hz update rate is not relevant; all damping
values are adjusted as described.
8.8 Configuring the update rate
The update rate is the rate at which the sensor reports the process variables to the
transmitter. This affects transmitter response time to changes in the process.
There are two settings for the update rate:
Normal and Special.
When
Normal is configured, all process variables are reported at the rate of
20 times per second (20 Hz).
When
Special is configured, a single, user-specified process variable is
updated at 100 times per second (100 Hz), and all other process variables are
updated at 6.25 times per second (6.25 Hz). If you set the update rate to
Special, you must also specify which process variable will be updated at
100 Hz.
Table 8-5 Valid damping values
Process variable Update rate
(1)
(1) See Section 8.7.3.
Valid damping values
Flow (mass and volume) Normal (20 Hz) 0, .2, .4, .8, ... 51.2
Special (100 Hz) 0, .04, .08, .16, ... 10.24
Density Normal (20 Hz) 0, .2, .4, .8, ... 51.2
Special (100 Hz) 0, .04, .08, .16, ... 10.24
Temperature Not applicable 0, .6, 1.2, 2.4, 4.8, ... 76.8
Model 1500 AN
Model 1700 AN
Model 1700 IS
Model 2500 CIO
Model 2700 AN
Model 2700 IS
Model 2700 CIO