
164 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Calibrating the Transmitter continued
C.4 Temperature calibration
Temperature calibration is a two-part procedure: temperature offset calibration and
temperature slope calibration. The entire procedure must be completed without
interruption.
You can calibrate for temperature with ProLink II software. See Figure C-5.
Figure C-5 Temperature calibration – ProLink II
• Model 1500 AN
• Model 1700 AN
• Model 1700 IS
• Model 2500 CIO
• Model 2700 AN
• Model 2700 IS
• Model 2700 CIO
Fill sensor with
low-temperature fluid
Fill sensor with
high-temperature fluid
Calibration
Temp slope cal
Do Cal
Enter temperature of
high-temperature fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Wait until sensor achieves
thermal equilibrium
Calibration
Temp offset cal
Do Cal
Enter temperature of
low-temperature fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Wait until sensor achieves
thermal equilibrium
Close Close
Komentarze do niniejszej Instrukcji