Emerson MICRO MOTION 2700 Instrukcja Instalacji Strona 139

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 138
Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 129
Troubleshooting continued
Troubleshooting DiagramsDefaultsCompensation
10.18 Setting the HART polling address to zero
If the transmitter’s HART polling address is set to a nonzero number, the primary mA output is fixed
at 4 mA. In this situation:
The primary mA output will not report process variable data.
The primary mA output will not indicate fault conditions.
If the HART address is set to zero, the primary mA output will report the primary variable on a
4–20 mA scale.
See Section 8.15.2.
10.19 Checking the flow measurement unit
Using an incorrect flow measurement unit can cause the transmitter to produce unexpected output
levels, with unpredictable effects on the process. Make sure that the configured flow measurement
unit is correct. Check the abbreviations; for example, g/min represents grams per minute, not gallons
per minute. See Section 6.4.
10.20 Checking the upper and lower range values
A saturated mA output or incorrect mA measurement could indicate a faulty URV or LRV. Verify that
the URV and LRV are correct and change them if necessary. See Section 6.5.2.
10.21 Checking the frequency output scale and method
A saturated frequency output or an incorrect frequency measurement could indicate a faulty frequency
output scale and/or method. Verify that the frequency output scale and method are correct and change
them if necessary. See Section 6.6.
10.22 Checking the characterization
A transmitter that is incorrectly characterized for its sensor might produce inaccurate output values. If
the flowmeter appears to be operating correctly but sends inaccurate output values, an incorrect
characterization could be the cause.
If you discover that any of the characterization data are wrong, perform a complete characterization.
See Section 6.2.
10.23 Checking the calibration
Improper calibration can cause the transmitter to send unexpected output values. If the transmitter
appears to be operating correctly but sends inaccurate output values, an improper calibration may be
the cause.
Micro Motion calibrates every transmitter at the factory. Therefore, you should suspect improper
calibration only if the transmitter has been calibrated after it was shipped from the factory.
The calibration procedures in this manual are designed for calibration to a regulatory standard. See
Appendix C. To calibrate for true accuracy, always use a measurement source that is more accurate
than the flowmeter. Contact the Micro Motion Customer Service Department for assistance.
Note: Micro Motion recommends using meter factors, rather than calibration, to prove the meter
against a regulatory standard or to correct measurement error. Contact Micro Motion before
calibrating your flowmeter. For information on meter factors, see Section 8.13.
Przeglądanie stron 138
1 2 ... 134 135 136 137 138 139 140 141 142 143 144 ... 263 264

Komentarze do niniejszej Instrukcji

Brak uwag