NOTE:
The Statistical Process Monitoring diagnostic capability in the Rosemount
3051S F
OUNDATION fieldbus Pressure Transmitter calculates and detects
significant changes in statistical parameters derived from the input process
variable. These statistical parameters relate to the variability of and the noise
signals present in the process variable. It is difficult to predict specifically
which noise sources may be present in a given measurement or control
application, the specific influence of those noise sources on the statistical
parameters, and the expected changes in the noise sources at any time.
Therefore, Rosemount cannot absolutely warrant or guarantee that Statistical
Process Monitoring will accurately detect each specific condition under all
circumstances.
SPM CONFIGURATION
AND OPERATION
The following section describes the process of configuring and using the
Statistical Process Monitoring diagnostic.
SPM Configuration for
Monitoring Pressure
Most Advanced Diagnostics Applications require using the device’s pressure
measurement as the SPM input. To configure the first SPM Block (SPM1) to
monitor the pressure set the following parameters:
SPM1_Block_Tag = TRANSDUCER
NOTE:
By default, as shipped from the factory, the tag of the sensor transducer block
is “TRANSDUCER”. DeltaV does not change the transducer block tags when
the device is installed and commissioned. However, it is possible that other
Fieldbus host systems may change the transducer block tags. If this happens,
SPM#_Block_Tag must be set to whatever tag was assigned by the host.
SPM1_Block_Type = TRANSDUCER Block
SPM1_Parameter_Index = Pressure (inH
2
O @ 68 °F)
SPM1_User_Command = Learn
(optional) SPM_Monitoring_Cycle = [1 – 5] minutes (see “Other SPM
Settings” on page 6-7)
(optional) SPM_Bypass_Verification = [Yes/No] (see page 6-7)
Apply all of these above changes to the device. Finally, set
SPM_Active = Enabled with 1st-order HP Filter
After SPM is enabled, it will spend the first 5 (whatever the
SPM_Monitoring_Cycle is set to) minutes in the learning phase, and then
another 5 minutes in the verification phase. If a steady process is detected at
the end of the verification phase, the SPM will move into the monitoring
phase. After 5 minutes in the monitoring phase, SPM will have the current
statistical values (e.g. current mean and standard deviation), and will begin
comparing them against the baseline values to determine if an SPM Alert is
detected.